品牌
生产厂家厂商性质
所在地
二氧化氮检测管9D NO2 Nitrogen dioxide
面议硫化氢检测管4D H2S Hydrogen sulphide
面议氟化氢检测管17D HF Hydrogen fluoride
面议过氧化氢检测管32D Hydrogen peroxide
面议乙醇检测管112D C2H5OH Ethanol
面议三氯乙烯检测管132D Trichloroethylene Trichloroethylene
面议一氧化碳气体检测管1D CO Carbon monoxide
面议二氧化碳检测管2D Carbon dioxide CO2
面议1,3 -丁二烯检测管174D 1,3-Butadiene
面议甲苯器检测管122DL C6H5CH3 Toluene
面议氨水氨气检测管3DL NH3 Ammonia
面议氨气气体检测管3D NH3 Ammonia
面议
Performance |
Measuring Range | 3 to 300 ppm |
Sampling Hours | 1 to 8 hours |
Detecting Limit |
|
Color Change | Yellow→ Purple |
Reaction Principle | Cl2C:CHCl +PbO2 +H2SO4→ HCl |
Coefficient ofVariation | 10%(for 25 to 300 ppm-hr) |
ShelfLife | 1 Year |
Correctionsfor temperature & humidity | Temperature correction isnecessary |
Store the tubes inthe refrigerator to keep at 10ºC (50ºF) orbelow. |
Possible coexisting substances and theirinterferences | (NOTE) |
Substance | Concentration | Interference | Change color by itself |
Hydrogen chloride,Chlorine | + | Purple | |
1,2-Dichloroethylene,Tetrachloroethylene | + | Purple | |
Toluene, Xylene | No | No |
Othersubstance measurable with this detectortube |
Substance | Correction Factor | Sampling Time | Measuring Range |
trans-1,2-Dichloroethylene | 2.0 | 1 to 8 hours | 6 to 600 ppm |
Chlorine | 0.8 | 1 to 8 hours | 2.4 to 240 ppm |
Hydrogen chloride | 0.6 | 1 to 8 hours | 1.8 to 180 ppm |
Tetrachloroethylene | 0.5 | 1 to 8 hours | 1.5 to 150ppm |
Calibration gasgeneration | Diffusion tubemethod |
TLV-TWA | TLV-S | Explosive range |
10 ppm | 25ppm | - |